System Configuration

Saliva testing can be conducted easily with this simple system.

The ST-4910 is used in combination with the PC and printer in the clinic. Both Windows and Mac are supported.

The SiLL-Ha is used in combination with the PC and printer in the clinic.
Both Windows and Mac are supported.


Test kit

Test kit

Test kit includes test strips and oral rinse solutions (3ml).

Test kit includes test strips and oral rinse solutions (3ml).

One test kit box contains 30 test strips, 30 packs of oral rinse solutions (3ml each) and a package insert.
*Dropper and cups are not included.

This system is not designed as a medical instrument or in vitro diagnostic product.
Therefore, the test result cannot be used for the purpose of determining disease or assisting in making diagnosis.

 

Specifications

Specifications

Instrument Specifications

NameSalivary Testing Instrument SiLL-Ha
ConfigurationInstrument, external PC software (Salivary Test Result Sheet Creator)
Measurement objectMeasurement object
Test stripSiLL-Ha paper
Measurement itemsCariogenic bacteria, acidity, buffer capacity, blood,leukocyte, protein, ammonia
Measurement principleDual wavelength reflectometry
Measurement wavelength565/635/760 (nm)
Sample supplyManual dropping
Measurement timeApprox. 5min/measurement sample
Number of measurement samples1 measurement sample
Warm-up timeMaximum 30sec
External outputConnectable with PC using USB port
Measurement environmentTemperature:10-30℃,Humidity: 30~60% RH (no condensation)
Storage environmentTemperature:1~35℃,Humidity:20~80% RH (no condensation)
Environment during transport Temperature:-10~60℃,Humidity:20~80% RH (no condensation)
DimensionApprox. 160 (W) 105 (D) 51 (H) mm
WeightApprox. 430g (including a rechargeable battery)
Power supplyUSB (5VDC, rechargeable battery used)
Power supply DC2.5W or under
Sound pressure level80dB or under
Location of useFor indoor use only
Maximum altitude Up to 2000m
Pollution degree2
Expected life5 years (According to company data)

PC Hardware Requirements

OS Windows*7 Japanese version (32 bit ver. / 64 bit ver.)
Processor Intel Pentium II 266mHz or above
Main memoryFor Windows*7/8/8.1/10: 1GB or above
For Mac OSX: 2GB or above
HDD/SDD available space1GB or above
ResolutionVGA (640x480) or above

Measurement Principle

This instrument measures samples (oral rinse sample) using the dual-wavelength reflectance method. The change in color of the test strip after sample application is measured by the reflectance of light with wavelengths of 565nm, 635nm and 760nm. Then, results are obtained for cariogenic bacteria, acidity, buffer capacity, leukocyte, protein and ammonia.